Maximising the Ion-Chamber reading

The usual ion chamber reading, measured through ion chamber 2 with slits set at 200 x 200 microns and with the SRS running at 200 mA, is between 20 and 30 on gain x1.

The detector system on Station 9.6 has its own locally built table. It has to be driven in the vertical and horizontal directions in order to maximise the alignment between the synchrotron beam and the internal collimation system of the detector being used. There are several motors used to do this, all of which are controlled through the beam alignment software pxgen++.

There are two ion chambers, supplied by MAR, that produce a reading on the black display terminal in the User area outside the hutch. CHAMBER1 (IC1) is situated immediately after the first set of slits on the camera, and is on the input side, closer to the source. Similarly, CHAMBER2 (IC2) is situated after the second set of slits, on the crystal side. If the display shows no reading, check that the port and station shutter are both open.

Beam Alignment: The PXGEN++ Program

The program may be started by typing pxgen++ at the prompt. Select the "Optimize Beam" tab.

Sample Table

A schematic of the table is shown. It is possible to move the following motors:

oemove1

A motor is selected by clicking on the appropriate arrow on the table schematic. To move the motor, first select the mode By with an increment of 50, units of Microns and Slow speed. The motor can then be moved by clicking on the + or the - button or by clicking on the Start Move button. It may be a good idea to record the motor positions at the beginning. Thus it will be possible to get back to the 'starting point' should some disaster occur!

It is recommended that you start by optimising the reading of IC1 by moving just the front vertical and front horizontal motors. When this value has been optimised, switch to IC2 and optimise this using the Yaw and Pitch options. These movements will maintain the reading of IC1 and allow IC2 to be optimised. An increment value of 10 to 20 mDeg should be used for Yaw and Pitch.

Reduction of the Cross-fire of the Primary Beam

Moving the horizontal pre-mono slits will change the horizontal crossfire of the beam hitting the monochromator, also changing the intensities on the ion chambers proportionally. Reduction of the crossfire will allow better resolution of spots in the horizontal direction. This is probably only needed for crystals with very large cell dimensions. When reducing the cross-fire, the slits should clip the beam evenly on both sides, i.e., one slit should reduce intensity by 25%, for example, and the other slit by another 25% of the original reading. Please remember to make a note of any changes to these slits in the log book.

oemove3

Beam Alignment: Trouble Shooting


Home, Next, Up